1. Cov neeg ua haujlwm yog siv cov khoom siv tshuaj lom neeg kev nyab xeeb tsis raug, nws cov kev tiv thaiv yuav cuam tshuam zoo heev. Tus neeg teb xov tooj yuav tsum tau ceev faj kom tswj tau qhov kev ncaj ncees ntawm cov huab cua ntawm lub hauv ntej qhib thaum txav caj npab mus rau hauv thiab tawm ntawm lub txee, thiab caj npab yuav tsum maj mam nkag mus thiab tawm ntawm lub hauv ntej qhib vertically.
Insert hands and arms into the biological safety cabinet and wait about a minute for the cabinet to adjust and allow the air inside to "sweep" over the surfaces of the hands and arms before proceeding with the item. Place all necessary items in the safety cabinet before starting the experiment to minimize the number of times the arms are in and out of the front opening.

2. Item placement The air intake grill in front of the secondary biological safety cabinet cannot be blocked by paper, equipment or other items. Items placed in the safety cabinet should be decontaminated with 70 percent alcohol. Experiments can be performed on a towel soaked in disinfectant to absorb droplets that may spill. All items should be placed as close to the edge of the table as possible at the back of the table and so that they do not block the alternate grille during operation. Equipment that can generate aerosols (eg mixers, centrifuges, etc.) should be placed near the rear of the safety cabinet. Larger items such as biohazardous waste bags, trays for discarded straws, and suction filter bottles should be placed on one side of the cabinet. Laboratory operations on the work surface should be carried out in the direction from the clean area to the contaminated area. Autoclavable biohazardous waste bags and straw trays should not be placed outside the cabinet, as these items must be used with the arms in and out of the cabinet frequently, which can interfere with the integrity of the cabinet's air barrier. This affects the protection of people and objects.
3. Operation and Maintenance Most biological safety cabinets are designed to work 24 hours a day. The researchers also found that continuous work helped control dust and particle levels in the lab. Class A1 and A2 biological safety cabinets that exhaust air into the room or are connected to special exhaust ducts through casing connections can be closed when not in use. Others, such as Class II B1 and Class II B2 biological safety cabinets, are installed through hard ducts and must maintain air flow at all times to maintain room air balance. Before starting the work and after finishing the work, let the safety cabinet work at least 5 minutes to complete the "purification" process, and should leave the time when the polluted air is sent out of the safety cabinet. All maintenance work on biological safety cabinets should be performed by qualified personnel. Any failures in the operation of the biological safety cabinet should be reported and repaired before reuse.
4. UV Teeb UV teeb tsis tas yuav tsum muaj nyob rau hauv cov khoom siv roj ntsha. Yog tias siv UV teeb, lawv yuav tsum tau ntxuav txhua lub lim tiam kom tshem tawm cov plua plav thiab av uas tuaj yeem cuam tshuam rau lawv cov kab mob zoo. Thaum lub txee tau lees paub, qhov siv ntawm UV teeb raug kuaj xyuas kom ntseeg tau tias lub teeb pom kev zoo. Nco ntsoov tua lub teeb UV thaum muaj neeg nyob hauv chav los tiv thaiv qhov muag thiab tawv nqaij los ntawm kev raug mob.
5. Cov nplaim taws qhib Kev siv cov nplaim taws yuav tsum zam kom tsis txhob nyob hauv ib puag ncig uas muaj cov kab mob me me uas tsim nyob rau hauv lub txee kev nyab xeeb. Kev siv cov nplaim taws tuaj yeem cuam tshuam rau huab cua thiab tuaj yeem ua rau muaj kev phom sij thaum tuav cov khoom tsis zoo thiab cov nplaim taws. Thaum sterilizing inoculation loops, lub micro burner los yog lub qhov cub hluav taws xob yuav tsum tau siv es tsis txhob qhib nplaim taws.

6. Spills Laboratory txoj kev khiav hauj lwm txoj cai yuav ua li cas los daws cov spills yog muab tso rau hauv lub chaw soj nstuam thiab yuav tsum tau nyeem thiab to taub los ntawm txhua tus neeg uas siv lub chaw soj nstuam. Nyob rau hauv qhov xwm txheej ntawm cov khoom uas muaj kev puas tsuaj lom nyob rau hauv lub txee rau kev nyab xeeb, nws yuav tsum tau muab ntxuav tam sim ntawd thaum lub txee ua haujlwm. Siv cov tshuaj tua kab mob kom zoo thiab txo qis aerosol tiam thaum tuav. Tag nrho cov ntaub ntawv uas tuaj rau hauv kev sib cuag nrog cov nchuav yuav tsum tau sterilized thiab / los yog autoclaved.
7. Certification After installation at regular intervals, each biological safety cabinet should be certified for operational performance and integrity in accordance with the manufacturer's instructions by qualified personnel to check that it meets national and national performance standards. The evaluation of the protective effect of the safety cabinet should include the integrity of the safety cabinet, the leakage of the HEPA filter, the velocity of the downdraft, the velocity of the frontal airflow, the number of negative pressure/air changes, the smoke pattern of the airflow, and the alarm and interlock system. test. You can also choose to test for leakage, illuminance, UV intensity, noise level and vibration. When conducting these tests, the inspectors should undergo specialized training and use specialized techniques and equipment. Testing by qualified personnel is strongly recommended.
8. Cleaning and Disinfection Since the remaining medium may allow microorganisms to grow and multiply, at the end of the experiment, all items in the biological safety cabinet, including equipment, should be cleared of surface contamination and removed from the safety cabinet. Before and after each use, decontaminate the inner surfaces of the biological safety cabinet. Work surfaces and interior walls should be wiped down with a disinfectant that kills any microorganisms that may be found in the cabinet. At the end of each day's test, the work surface of the biological safety cabinet, the surrounding area, and the inside and outside of the glass should be wiped to remove surface contamination. When effective against the target organism, a bleach solution or 70 percent alcohol can be used for disinfection. After using aggressive disinfectants such as bleach, it must be wiped again with sterile water. It is recommended to keep the safety cabinet in operation. If shutting down, it should run for 5 minutes before shutting down to purge the gas inside.
9. Decontamination Biosafety txee yuav tsum tau decontaminated ua ntej tshem tawm thiab lim hloov. Txoj kev siv feem ntau yog siv formaldehyde vapor fumigation. Decontamination ntawm cov khoom siv tshuaj lom neeg kev nyab xeeb yuav tsum tau ua los ntawm cov neeg ua haujlwm tsim nyog.
10. Personal Protective Equipment Personal protective clothing should be worn when using the biological safety cabinet. Ordinary test suits may be worn during operations at Biosafety Levels 1 and 2. Reverse-back test gowns with front reinforcement provide better protection and should be used when performing biosafety level III and IV operations (except for gown-type laboratories). Gloves should be worn over the gown, and elasticated sleeves may be worn to protect the researcher's wrists. Masks and safety glasses may also be required for some operations.
11. Siren Lub txee kev nyab xeeb lom neeg tuaj yeem nruab nrog kev xaiv ntawm ob hom sirens.